Relevant parameters for laser surgery of soft tissue | Scientific Reports

News

HomeHome / News / Relevant parameters for laser surgery of soft tissue | Scientific Reports

Oct 14, 2024

Relevant parameters for laser surgery of soft tissue | Scientific Reports

Scientific Reports volume 14, Article number: 1263 (2024) Cite this article 1181 Accesses Metrics details In recent years, the laser has become an important tool in hospitals. Laser surgery in

Scientific Reports volume 14, Article number: 1263 (2024) Cite this article

1181 Accesses

Metrics details

In recent years, the laser has become an important tool in hospitals. Laser surgery in particular has many advantages. However, there is still a lack of the understanding of the influence of the relevant parameters for laser surgery. In order to fill this gap, the parameters pulse frequency, use of an exhaustion system, air cooling, laser power, laser scan speed, laser line energy and waiting time between cuts were analysed by ANOVA using inter-animal variation as a benchmark. The quality of the cuts was quantized by a previously published scoring system. A total of 1710 cuts were performed with a \(\text {CO}_2\) laser. Of the parameters investigated, laser power and scan speed have the strongest influence. Only the right combination of these two parameters allows good results. Other effects, such as the use of pulsed or continuous wave (CW) laser operation, or air cooling, show a small or negligible influence. By modulating only the laser power and scan speed, an almost perfect cut can be achieved with a \(\text {CO}_2\) laser, regardless of the external cooling used or the laser pulse duration or repetition rate from CW to nanosecond pulses.

For many years, laser surgery has been an accepted tool in various surgical fields1. The use of lasers in hospitals is also increasing2. The advantage of lasers is that they can achieve results similar to those of conventional surgery while being minimally invasive3,4. At the same time, laser surgery usually has a high healing potential, with less post-operative inflammation and swelling5. The coagulation effect of the laser radiation improves visibility by coagulating small blood vessels6. While these benefits are widely accepted, the practical advantage is still debated. For example, Seifi and Matini7 showed in a small meta-study that there was no benefit from cutting soft dental tissue with a laser. They conclude: “Introducing an appropriate laser with suitable wavelength, input power and other properties for mentioned indications needs more research and clinical trials”.

This shows that there is still a lot of research to be done. Even in recent studies, the types of lasers used are still under investigation8. For a more general view, it is important to look more closely at the laser-material interaction to differentiate the regimes of material ablation. Boulnois9 distinguishes between vaporisation, photoablation and photodisruption. These mechanisms play an important role in the efficiency of laser surgery. For example, Werner et al.10 showed that a \(\text {CO}_2\) laser could achieve more than ten times the ablation rate of the frequency doubled \(\text {ND:YVO}_4\) laser at 532 nm.

Sánchez et al.11 compared a \(\text {CO}_2\) laser with an Er,CR:YSGG laser on gingiva. While the \(\text {CO}_2\) laser cut faster and without bleeding, the Er,CR:YSGG allowed a faster healing time. A slower healing ability of the \(\text {CO}_2\) laser was also found in later studies12. However, a meta-analysis from 201913 shows that a subgroup analysis for the type of laser cannot be done because the amount of data is too small. Furthermore, Protásio et al.13 conclude “... that labial frenectomies performed with high-intensity surgical lasers are faster and offer a better prognosis in terms of pain and discomfort during speech and chewing than those performed with conventional scalpels”, if publication bias is not taken into account. A later randomized double-blinded and controlled pediatric clinical study in 2021 by Fioravanti et al.14 with obstructive sleep apnea syndrome (OSAS) was performed. There, a milli second pulsed laser with a wavelength of 980 nm strongly decreased the OSAS compared to the control group contrasting the result from the review by Protásio et al.13. Also, a case report in 2023 showed a good outcome for healing of osteonecrosis of the jaw15 by laser treatment with a 980 nm laser. Moreover, a review from Lesniewski et al.16 reveals that “... diode lasers and LEDs are equally effective tools for the phototherapy in periodontology and oral surgery”. Therefore, currently no firm conclusion can be drawn regarding whether laser surgery and which laser type result in a better surgery performance in the oral cavity. However, lasers in the near infrared range seem to be advantageous16.

While this was an example of labial frenectomies, one of the main drawbacks of some studies is the statistics with too few samples and examined parameters. For example, the highly cited study by Cercadillo-Ibarguren et al.17 varied the four parameters laser power, laser type, air spray and pulsed laser operation with 117 samples. It was concluded that Er,CR:YSGG lasers performed well, while diode and \(\text {CO}_2\) lasers did not perform well. However, other authors state that a \(\text {CO}_2\) laser rarely causes any unwanted tissue damage when used correctly18,19,20.

While this is obviously a contradiction, the important question is: why are there differences in the results? To investigate this question, various parameters are collected from the literature. A study by El-Sherif and King21 for a laser with a wavelength of 2 \(\upmu\)m showed that the pulsed mode results in less damage of soft tissue than the CW mode of the laser. In addition, the heat affected zone ranges from 120 to 160 \(\upmu\)m for the pulsed laser mode and 400–800 \(\upmu\)m for the CW mode21. For bone tissue, the heat affected zone is for a \(\text {CO}_2\) laser only 6 \(\upmu\)m in comparison22. Similar results were found for an Er:YAG laser with a wavelength of 2.94 \(\upmu\)m. While the heat affected zone for the q-switched laser was 5–10 \(\upmu\)m, it increased to 10–50 \(\upmu\)m for the spiking mode with longer pulse durations23. Furthermore, the results did not change for different tissue types including soft and hard tissue23. It can also be concluded from the results of Krapchev et al.24 that the pulse frequency and duration should be below the thermal relaxation time to prevent unwanted tissue damage. This leads to the first two potential influencing parameters: wavelength and pulse duration.

Another factor to consider is cooling of the tissue. Ivanenko et al.22 were able to show that small amounts of water prevented carbonisation, while larger amounts of water showed no further improvement. This is due to the fact that the water evaporates when too much heat is present, leading to a cooling effect25. More specifically, the additional water creates a temperature gradient from the water to the tissue, causing the heat to flow to the water rather than the surrounding tissue26. For air cooling, the results are contradictory: while Ivanenko and Hering27 claim that pure gas cooling leads to more thermal damage, Afilal28 claims the opposite. This leads to the next two potential influencing parameters: water cooling and air cooling.

Finally, the laser parameters should be considered. While the laser power is an obvious influencing parameter, the effect of scan speed could be shown by Afilal28. He was able to show that carbonisation could be prevented by using the correct scanning parameters. In another case, the temperature increase could be limited to 30 K instead of 400 K by using a scanning technology29. It is also known from the field of laser material processing (e.g. welding, cutting and additive manufacturing with lasers) that the line energy is an important parameter. This leads to the final potential influencing parameters: laser power, laser scan speed and laser line energy.

In total, seven potential influencing parameters that affect the laser surgery process can be identified from the literature: wavelength, pulse duration, water cooling, air cooling, laser power, laser scan speed and laser line energy. To quantify the quality of the cut, the previously published scoring system30 was used. With the help of the scoring system, all previously mentioned parameters, except wavelength and type of cooling, will be investigated in this study because studying the influence of wavelength would require too many different lasers and water cooling is known to result in a significant improvement in cut quality at the cost of increased complexity31,32 and longer cutting times. In addition, the number of breaks between laser scans and the effect of an exhaust system are studied. The number of breaks is used to study the effect of heat dissipation on the cut. The exhaust parameter is taken into account as it may be important to remove the plumes from the laser surgery process. All these parameters are compared to the effect of the inter animal variation as a benchmark for their importance.

The method section consists of four sections. In the first two sections, experimental procedure including the origin of the tissue and the statistical analysis is explained. In the following two sections, the varied parameters are discussed in detail.

Setup with tissue, the compressed air supply, the exhaustion system and the laser source.

In this section, the set-up and the experimental parameters that are not altered in this study are explained. The experimental setup is shown in Fig. 1. The used laser is from MICROSTORM (FEHA LasterTec GmbH, Germany) with a wavelength of 10.6 \(\upmu\)m. It is a diffusion cooled \(\text {CO}_2\)-laser with an acousto-optical modulator. The modulator is driven by an external pulse generator (DG1032Z, Rigol Technologies).

The experimental parameters are divided into constant and variable parameters. In this section only the constant parameters are explained and summarized in Table 1. The temporal shape of the laser is rectangular. The focal distance is 13.7 cm from the lens to the tissue sample. The focal diameter is 258 \(\upmu\)m. Furthermore, the deflection of the laser beam is realised by a scan head (Scanlab GmbH). This allows the laser beam to be moved at a controlled speed of the sample. The temporal laser profile is set to be rectangular to ensure that the same light intensity is always applied to the tissue surface. A total of 1 cm long incisions were cut. There are 20 scans for each incision with a scan time of 10 ms. This leads to a cutting depth of up to 4 mm. If breaks are made between scans, the break time is 2 s. If the number of breaks is one, a break is done after 10 scans, if it is 3, a break is done after each 5 scans and if the number is 19 a break is done after each scan. The laser comes perpendicular to the surface of the table, However in practice, the laser may not be entirely perpendicular to the tissue as it is not flat. This can be seen in Fig. 1. Hence, the incident angle of the laser is seen as to be approximately zero degree.

For all experiments, bisected pork tissues of food quality were purchased from the local butcher. Therefore, an ethical proposal for animal experimentation is not required. The samples used were 1 cm thick pieces of pork from the topside. The laser power of 235 W for 0.2 s produced the highest energy input observed in all experiments − 47 J. As a result, the sample size was always sufficiently large compared to the heat affected zone so that the size of the tissue had no effect on the heat dissipation.

Table 2 provides an overview about the number of samples. There are five repetitions for each animal sample and parameter setting for the non-laser parameters, resulting in a total of 30 data points (n = 30) per parameter setting. The experiments for the laser parameters include 3 different animal samples with 10 repetitions each, so there are 30 data points (n = 30) per parameter setting.

For the quantitative analysis all cuts scored on the scoring system presented in a previous paper30 from our group (especially Fig. 2 provides an good overview). In short, by looking at tissue damage at the rim or the cutting front of the cut, a score from 1 to 5 is given, where 5 is the best possible score. The scoring is performed based on the presence/amount of carbonization and the colour of the tissue. Figure 2 shows examplary scores for different cuts. It should be noted that a score of 3–4 already denotes a pretty good cut in comparison to cuts from actual procedures such as in Fig. 4b from Vanderhem et al.33. The cut shown there would be scored as 2. As it could be shown that the scoring of the cutting front is more reliable30, the complete analysis in the study is based on the scores of the cutting front.

Exemplary scoring for the five different cuts (taken from our previous publication30).

For each data set, the effect of the variable is presented and the samples are compared for significant difference. All commands are taken from SciPy34 and the names in parentheses indicate the corresponding SciPy command. The difference between the samples is tested with the Wilcoxon–Mann–Whitney-Test (“scipy.stats.mannwhitneyu”) as the scoring is a ordinal scale. As multiple comparisons are performed, the significance levels are reduced to 0.01 (*), 0.001 (**), 0.0001 (***) and 0.00001 (****). Afterwards, an analysis of variance (ANOVA) is performed with the help of the statsmodel framework35.

The aim of this part is to evaluate the influencing parameters considered relevant in the literature, except the laser parameters. An overview of the varied parameters is given in Table 3. In order to reduce the number of experiments required, two sets of experiments have to be performed: In experiment A, the effect of air cooling, number of breaks and pulse duration is investigated. In experiment B, the effect of exhausting the plumes is investigated. In both data sets, each parameter combination is repeated five times on six different animals. To conduct the analysis, a benchmark is required to determine the relevance of a given parameter. This can be established by examining the variance in the effects of cutting different animals as for a practical applications, the effect of cutting different animals has to be lower than the effect of a given parameter. By attributing how much variance of the different animals is affecting the scoring, a benchmark is generated that takes into account the tissue’s storage and pre-treatment.

The experiments in experiment A are performed with two laser powers (P): \(P=83\) W (irradiance = 160 \(\text {kW}/\text {cm}^2\)) and P = 129 W (irradiance = 247 \(\text {kW}/\text {cm}^2\)), in order to evaluate all non-laser parameters for a more optimal case (83 W) and a case with more tissue damage (129 W). For the first intensity, it is tested if the parameters can worsen the cut and in the second case, if the tested parameters can lead to optimal laser cuts. For all experiments, the scan speed is kept constant at 1 \(\text {m}/\text {s}\), resulting in an illumination time of 10 ms. Furthermore, for both laser powers a separate ANOVA analysis is performed.

For experiment B, only the laser power of 129 W and the pulse duration of 964 ns are used. The parameters number of breaks and air cooling are varied as in A. The number of parameters is reduced to decrease the number of experiments required. The pulse duration is excluded as it is a laser parameter, and the laser power of 129 W is chosen to evaluate if the exhaustion of the plumes can lead to an improvement of a non-optimal laser cut.

The aim of this part is to evaluate the influencing laser parameters. An overview of the parameters is provided in Table 4. For parameters, the line energies (\(E_l\)) are chosen to be: 83 \(\text {J}/\text {m}\), 172 \(\text {J}/\text {m}\) and 235 \(\text {J}/\text {m}\). The line energy is calculated as followed:

where \(v_{scan}\) is the scan speed. A total of 270 cuts are evaluated for three animals: 90 for each line energy. For each line energy, 3 laser powers/scanning speeds are selected. Thus, each parameter combination is performed ten times.

Each section of the results is divided into two parts. The first part presents the means and compares the significance of the data sets. The second part presents the results of the ANOVA analysis.

Effect of the parameters number of breaks, pulse duration and cooling by air for the laser power of 83 W. The errorbars show the standard deviation. The significance levels are 0.01 (*), 0.001 (**), 0.0001 (***) and 0.00001 (****) by a Wilcoxon–Mann–Whitney-test.

Figures 3 and 4 show the average effect and the different parameters with standard deviation and significance levels. For the laser power of 83 W, the cutting quality is overall higher as the cuts with a laser power of 129 W. This is due to the fact that the laser power of the cuts with 129 W is chosen to be too high to see the possible improvements by the different parameters. In both cases, a higher number of breaks reduces the unwanted tissue damage. Also for both pump powers, air cooling and different pulse durations show no or little significance and variation.

For 83 W, a shorter pulse duration seems to reduce the cutting quality. However, the significance is small. The air pressure seems to reduce the cutting quality, but it is not a significant influence. The already good results could be improved by giving the laser more breaks.

Effect of the parameters number of breaks, pulse duration and cooling by air for the laser power of 129 W. The errorbars show the standard deviation. The significance levels are 0.01 (*), 0.001 (**), 0.0001 (***) and 0.00001 (****) by a Wilcoxon–Mann–Whitney-test.

For 129 W, there is no significant effect of pulse duration. As for the 83 W, the air pressure seems to reduce the cutting quality a little. This effect is slightly significant. As the laser power was too high, the increase in the number of breaks shows a strong increase in the cutting quality. This could be due to the fact that the heat has more time to diffuse into the surrounding tissue, resulting in less heat accumulation. As more cuts reach five points, even the coagulated tissue is decreased.

Tables 5 and 6 show the results of the ANOVA for a laser power of 83 and 129 W, respectively. Overall, the results are similar to the previous analysis. The effect of breaks is comparably large and highly significant. In addition, the effect of inter-animal variation is highly significant and can explain more of the variance than pulse duration and air cooling. Therefore, in addition to the conflicting significance, the effect of these two parameters can be discarded as they explain less variance. It should also be noted that \(R^2\) and especially the adjusted \(R^2\) is relatively low. Thus, most of the variance cannot be explained.

For 83 W, air cooling shows no significant effect and the explainable variance is low. While the pulse duration shows a significant effect, the explainable variance is lower than for the inter-animal variation. Thus, the normal inter-animal variation masks the effect of both parameters. It should also be noted that the adjusted \(R^2\) is only 0.38.

For 129 W, the effect of the air cooling and the pulse duration are the opposite than for 83 W. The pulse duration shows no significant effect and the explainable variance is low. While air cooling shows a significant effect, the explainable variance is comparable to the inter-animal variation. For this case, the adjusted \(R^2\) is 0.45.

In summary, it can be concluded that, contrary to the literature, pulse duration has no or a contradictory effect on cutting quality. Furthermore, its effect on the final result is less than the inter-animal variation. Therefore, its effect can be ignored. Air cooling also has little or no effect on cutting quality. Therefore, air cooling should not be used, also in the interest of reducing the complexity of the set-up. The number of breaks between cuts shows a strong effect and significantly improves the results. In addition, the number of breaks leads to much stronger effects than the inter-animal variation. This parameter should therefore be taken into account. For practical application, however, there is a conflict of objectives. On the one hand, tissue damage should be low. This is favoured by a high number of breaks. On the other hand, the operation should be fast. This is favoured by a low number of breaks.

Effect of the parameters number of breaks, exhaustion and cooling by air for the laser power of 129 W. The errorbars show the standard deviation. The significance levels are 0.01 (*), 0.001 (**), 0.0001 (***) and 0.00001 (****) by a Wilcoxon–Mann–Whitney-test.

Figure 5 shows the average effect and the different parameters with standard deviation and significance levels. The results of the breaks and the cooling by air are almost identical to the results for 129 W in experiment A. Hence, the results of this study are reproducible. The plume exhaust parameter shows no significant effect. These results are also supported by the ANOVA in Table 7. The significance of the number of breaks is high and it explains a large amount of variance. At the same time, exhaust and air cooling have a minimal effect on the explainable variance and the significance level is much lower than for the number of breaks. Therefore, both air cooling and exhaust do not play a major role. Both parameters can be adjusted according to other requirements of laser surgery.

Effect of the line energy and the laser power/scan speed. The errorbars show the standard deviation. The significance levels are 0.01 (*), 0.001 (**), 0.0001 (***) and 0.00001 (****).

Figure 6 shows the effect of line energy and laser power or scan speed, respectively. It can be seen that the line energy has a strong influence on the resulting cut quality. The laser line energy at smallest value 83 \(\text {J}/\text {m}\), gives the best cutting quality. The scan speed of 1 \(\text {m}/\text {s}\) gives the best results regardless of the laser power and line energy. Thus, it appears that for a given line energy, the scan speed has a more significant effect than the laser power. For the comparable optimal line energy of 83 \(\text {J}/\text {m}\), the scan speed has only a small effect. For less optimal parameters, a non-optimal scan speed can further reduce the quality of the cut. The strongest effect of the scan speed occurs when the scan speed is low, whereas at higher scan speeds, the influence of the scan speed becomes insignificant. In summary, line energy is the most important parameter and scan speeds at 1 \(\text {m}/\text {s}\) or higher are preferred.

These overall results are supported by the ANOVA analysis which is shown in Table 8: all two laser parameters are significant and the line energy explains most of the variance. The explained variance is higher than for any of the non-laser parameters. As both parameters are also highly significant, it can be concluded that the laser parameters are the most important parameters. The effect of laser power/scan speed is analysed again separately for each line energy to clearly show whether laser power/scan speed or line energy is the more important parameter. It should be noted that the parameters of scan speed and laser power cannot be separated as they are indirectly proportional. The importance has to be concluded from Fig. 6.

Tables 9, 10 and 11 show the effect of laser power/scan speed for line energies of 83 \(\text {J}/\text {m}\), 172 \(\text {J}/\text {m}\) and 235 \(\text {J}/\text {m}\). For an optimal line energy of 83 \(\text {J}/\text {m}\) (Table 9), the effect of laser power/scanning speed is hardly significant and only 11% of the variance can be explained. This effect is supported by the low value of \(R^2\). Therefore, for an optimal line energy, laser power and scan speed are not important.

These results change when a non-optimal line energy is used, as shown in Tables 10 and 11. In this case, the effect of laser power and scan speed becomes significant and explains a higher percentage of the variance. The \(R^2\) value increases in this case. All three effects increase when the line energy is further away from the optimal value. Nevertheless, the optimal choice of laser power or scan speed can only improve the cut quality to a certain extent, which is limited by the non-optimal line energy. In other words: the line energy determines the maximum achievable cut quality.

The main limitation of this study is the fact that all experiments were performed in an ex-vivo setting. This clearly limits the generalisability of the results presented as important parameters such as time to heal cannot be assessed. However, the chosen ex-vivo setting allows the study of a large number of laser cuts and, due to the lack of perfusion, tissue damage may appear easier. Therefore, it is possible that parameters from the ex-vivo setting can be transferred to in-vivo.

The second limitation is that only one type of tissue is used for all laser cuts. While this is a requirement of this study to really understand the effect of different parameters, it does limit the generalisability. It is likely that the results can be transferred to at least some types of soft tissue, but generalisation to hard tissue such as bone is not possible without further experiments. However, with the current approach demonstrated, it may be possible to find a parameter setting that allows bone cutting with a \(\text {CO}_2\) laser.

A third limitation is that pig tissue is used instead of human tissue. Thus, the transferability is limited. However, the parameters tested show a much stronger effect than the inter-animal variation. Therefore, it is expected that the presented results are at least partially transferable to other species.

The fourth limitation is the size of the cuts and fixed laser focus. All cuts had a length of 1 cm and a depth of up to 4 mm. This leads to two conclusions. Deeper cuts are expected to result in more tissue damage. Therefore, the optimal parameters may be different or even have to be adjusted during laser surgery or in other words the optimal laser parameters might be depth dependent. Secondly, the fixed laser focus might have the effect of limiting the cutting depth. Hence, the cutting depth could be greater with an adjusted focus position.

In this study, the parameters of pulse duration, laser power, laser scan speed, line energy, air cooling, exhaust system and number of breaks between cuts were investigated. Among these parameters, line energy is the most important. It determines the maximum cutting quality that can be achieved. If the correct line energy is known, the scan speed and the number of breaks are similarly important parameters. Interestingly, the scan speed of about 1 \(\text {m}/\text {s}\) is optimal for all line energies tested. For the number of breaks, it can be said that the more breaks there are, the more time the heat has to dissipate and the less unwanted tissue damage there will be. However, the choice of the break numbers should consider the practical requirement. A lower number of breaks is required to speed up the laser surgery process. As the line energy has to be optimized and the scan should be around 1 \(\text {m}/\text {s}\), the laser power is fixed. All the other parameters (pulse duration, air cooling, exhaust system) are not relevant for the laser surgery with a \(\text {CO}_2\).

This leads to the following conclusions for the use of the \(\text {CO}_2\) laser in soft tissue. Because of the importance of line energy and scan speed, the \(\text {CO}_2\) laser is not suitable for a manual handheld device. The \(\text {CO}_2\) laser should be used with a robot or remote system to achieve high quality cuts. Nevertheless, it is possible to use a \(\text {CO}_2\) laser for tissue ablation. Therefore, the \(\text {CO}_2\) laser is still an attractive laser for some applications33,36,37. However, the \(\text {CO}_2\) laser shines under remote operation conditions. As the pulse duration of the \(\text {CO}_2\) laser has no effect on the cutting quality, any pulse duration can be used. However, this result cannot be transferred to near infrared laser types, as the \(\text {CO}_2\) laser is superficially absorbed, unlike e.g. Nd:YAG lasers operating at 1064 nm. However, for frequency doubled or tripled Nd:YAG lasers or Er:YAG lasers, the absorption is also comparably high which is caused by hemoglobin for the frequency doubled or tripled Nd:YAG lasers and water for the Er:YAG lasers. Hence, it makes sense to investigate if the pulse duration may be irrelevant for the cutting quality for these lasers. The small effect of the air cooling and exhaust system parameters can probably be generalised to most other laser types. Therefore, air cooling is not required and the exhaust system can be adapted to other requirements of the laser surgery system.

All data analysed during this study is included in the supplementary information files.

Spyropoulos, B. 50 years lasers: In vitro diagnostics, clinical applications and perspectives. Clin. Lab. 57(3–4), 131–142 (2011).

PubMed Google Scholar

Späth, M. et al. A quantitative evaluation of the use of medical lasers in German hospitals. J. Biophoton. 20, e201900238 (2019).

Google Scholar

Sinha, P. et al. Transoral laser microsurgery for oral squamous cell carcinoma: Oncologic outcomes and prognostic factors. Head Neck 36(3), 340–351 (2014).

Article PubMed Google Scholar

Thomson, P. J. & Wylie, J. Interventional laser surgery: An effective surgical and diagnostic tool in oral precancer management. Int. J. Oral Maxillofac. Surg. 31(2), 145–153 (2002).

Article CAS PubMed Google Scholar

Fisher, S. E. & Frame, J. W. The effects of the carbon dioxide surgical laser on oral tissues. Br. J. Oral Maxillofac. Surg. 22(6), 414–425 (1984).

Article CAS PubMed Google Scholar

Ben-Bassat, M., Kaplan, I., Shindel, Y. & Edlan, A. The co2 laser in surgery of the tongue. Br. J. Plast. Surg. 31(2), 155–156 (1978).

Article CAS PubMed Google Scholar

Seifi, M. & Matini, N.-S. Laser surgery of soft tissue in orthodontics: Review of the clinical trials. J. Lasers Med. Sci. 8(Suppl 1), S1 (2017).

Article PubMed PubMed Central Google Scholar

Doizi, S. et al. Comparison of holmium: Yag and thulium fiber lasers on soft tissue: An ex vivo study. J. Endourol. 36(2), 251–258 (2022).

Article PubMed Google Scholar

Boulnois, J.-L. Photophysical processes in recent medical laser developments: A review. Lasers Med. Sci. 1, 47–66 (1986).

Article Google Scholar

Werner, M., Klasing, M., Ivanenko, M., Harbecke, D., Steigerwald, H., & Hering, P. Co2 laser free form processing of hard tissue. In European Conference on Biomedical Optics, 6632_1 (Optica Publishing Group, 2007).

Pié-Sánchez, J., España-Tost, A. J., Arnabat-Domínguez, J. & Gay-Escoda, C. Comparative study of upper lip frenectomy with the co2 laser versus the er, cr: Ysgg laser. Med. Oral Patol. Oral Cirugia Fucal 17(2), e228 (2012).

Article Google Scholar

Sfasciotti, G. L., Zara, F., Vozza, I., Carocci, V., Ierardo, G, & Polimeni, A.,. Diode versus co2 laser therapy in the treatment of high labial frenulum attachment: A pilot randomized, double-blinded clinical trial. Int. J. Environ. Res. Public Health 17(21), 7708 (2020).

Article PubMed PubMed Central Google Scholar

Protásio, A. C. R., Galvão, E. L. & Falci, S. G. M. Laser techniques or scalpel incision for labial frenectomy: A meta-analysis. J. Maxillofac. Oral Surg. 18, 490–499 (2019).

Article PubMed PubMed Central Google Scholar

Fioravanti, M., Zara, F., Vozza, I., Polimeni, A. & Sfasciotti, G. L. The efficacy of lingual laser frenectomy in pediatric osas: A randomized double-blinded and controlled clinical study. Int. J. Environ. Res. Public Health 18(11), 6112 (2021).

Article PubMed PubMed Central Google Scholar

El Mobadder, M. et al. Management of medication-related osteonecrosis of the jaw with photobiomodulation and minimal surgical intervention. Dent. J. 11(5), 127 (2023).

Article Google Scholar

Lesniewski, A., Estrin, N. & Romanos, G. E. Comparing the use of diode lasers to light-emitting diode phototherapy in oral soft and hard tissue procedures: A literature review. Photobiomodul. Photomed. Laser Surg. 40(8), 522–531 (2022).

PubMed Google Scholar

Cercadillo-Ibarguren, I. et al. Histologic evaluation of thermal damage produced on soft tissues by co2, er, cr: Ysgg and diode lasers. Med. Oral Patol. Oral Cirugia Bucal 15(6), 912–918 (2010).

Article Google Scholar

Convissar, R. A. Laser biopsy artifact. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontol. 5(84), 458 (1997).

Article Google Scholar

Bornstein, M. M., Winzap-Kälin, C., Cochran, D. L. & Buser, D. The co 2 laser for excisional biopsies of oral lesions: A case series study. Int. J. Periodont. Restor. Dent. 25, 3 (2005).

Google Scholar

Tuncer, I., Özçakır-Tomruk, C., Şencift, K. & Çöloğlu, S. Comparison of conventional surgery and co2 laser on intraoral soft tissue pathologies and evaluation of the collateral thermal damage. Photomed. Laser Surg. 28(1), 75–79 (2010).

Article PubMed Google Scholar

El-Sherif, A. F. & King, T. A. Soft and hard tissue ablation with short-pulse high peak power and continuous thulium-silica fibre lasers. Lasers Med. Sci. 18(3), 139–147 (2003).

Article CAS PubMed Google Scholar

Ivanenko, M. M., Fahimi-Weber, S., Mitra, T., Wierich, W. & Hering, P. Bone tissue ablation with sub-\(\mu\)s pulses of a q-switch co 2 laser: Histological examination of thermal side effects. Lasers Med. Sci. 17, 258–264 (2002).

Article CAS PubMed Google Scholar

Walsh, J. R., Joseph, T., Flotte, T. J. & Deutsch, T. F. Er: Yag laser ablation of tissue: Effect of pulse duration and tissue type on thermal damage. Lasers Surg. Med. 9(4), 314–326 (1989).

Article PubMed Google Scholar

Krapchev, V. B, Rabii, C. D, & Harrington, J. A. Novel co2 laser system for hard tissue ablation. In Laser Surgery: Advanced Characterization, Therapeutics, and Systems IV, Vol. 2128, 341–348 (SPIE, 1994).

Zweig, A. D. et al. Lateral thermal damage along pulsed laser incisions. Lasers Surg. Med. 10(3), 262–274 (1990).

Article CAS PubMed Google Scholar

Spencer, P., Cobb, C. M., McCollum, M. H. & Wieliczka, D. M. The effects of co2 laser and nd: Yag with and without water/air surface cooling on tooth root structure: Correlation between ftir spectroscopy and histology. J. Periodontal Res. 31(7), 453–462 (1996).

Article CAS PubMed Google Scholar

Ivanenko, M. M. & Hering, P. Wet bone ablation with mechanically q-switched high-repetition-rate co2 laser. Appl. Phys. B 67, 395–397 (1998).

Article ADS CAS Google Scholar

Afilal, S. Ablationsmechanismen von biologischem Hartgewebe bei Bestrahlung mit kurzgepulsten CO2-Lasern. PhD thesis (2004).

Brendemühl, A., Werner, M., Ivanenko, M., Hering, P., & Buzug, T. M. Comparison of process temperature during laser and mechanical cutting of compact bone. In Advances in Medical Engineering, 419–424 (Springer, 2007).

Hohmann, M. et al. Development and evaluation of a scoring system for assessing incisions in laser surgery. Sci. Rep. 12(1), 14741 (2022).

Article ADS CAS PubMed PubMed Central Google Scholar

Bernal, L. M. B. et al. Optimizing deep bone ablation by means of a microsecond er: Yag laser and a novel water microjet irrigation system. Biomed. Opt. Express 11(12), 7253–7272 (2020).

Article CAS Google Scholar

Cetin, C. et al. Bone ablation using a ho: Yag laser. Curr. Direct. Biomed. Eng. 8(2), 580–583 (2022).

Article Google Scholar

Van der Hem, P. S., Nauta, J. M., Van der Wal, J. E. & Roodenburg, J. L. N. The results of co2 laser surgery in patients with oral leukoplakia: A 25 year follow up. Oral Oncol. 41(1), 31–37 (2005).

Article PubMed Google Scholar

Virtanen, P. et al. SciPy 1.0. fundamental algorithms for scientific computing in Python. Nat. Methods 17, 261–272 (2020).

Article CAS PubMed PubMed Central Google Scholar

Seabold, S., & Perktold, J. statsmodels: Econometric and statistical modeling with python. In 9th Python in Science Conference (2010).

Pecaro, B. C. & Garehime, W. J. The co2 laser in oral and maxillofacial surgery. J. Oral Maxillofac. Surg. 41(11), 725–728 (1983).

Article CAS PubMed Google Scholar

Kazakova, R. T. et al. Histological gingival assessment after conventional and laser gingivectomy. Folia Med. (Plovdiv) 60(4), 610–616 (2018).

Article PubMed Google Scholar

Download references

The authors gratefully acknowledge funding of the Erlangen Graduate School in Advanced Optical Technologies (SAOT) by the Bavarian State Ministry for Science and Art.

In the writing process during the preparation of this work, the authors used “DeepL Write” in order to improve language and readability. After using this tool, the authors reviewed and edited the content as necessary and they take full responsibility for the content of the publication.

Open Access funding enabled and organized by Projekt DEAL.

Institute of Photonic Technologies (LPT), Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Konrad-Zuse-Straße 3/5, 91052, Erlangen, Germany

Martin Hohmann, David Kühn, Dongqin Ni, Moritz Späth, Anindya Ghosh, Florian Stelzle, Florian Klämpfl & Michael Schmidt

Erlangen Graduate School in Advanced Optical Technologies (SAOT), Paul-Gordon-Straße 6, 91052, Erlangen, Germany

Martin Hohmann, Dongqin Ni, Moritz Späth, Anindya Ghosh, Maximilian Rohde, Florian Stelzle, Florian Klämpfl & Michael Schmidt

Department of Oral and Maxillofacial Surgery, University Hospital Erlangen, 91052, Erlangen, Germany

Maximilian Rohde & Florian Stelzle

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

M.H. conceptualized the research, did the data analysis, prepared the manuscript and supported D.K. in the lab work. D.K. did the experimental work did part of the data analysis. D.N., M.S. and A.G. supported the data analysis. M.R. specified the work from the medical point of view. F.S., F.K. and M.S. guided the general research strategy. All authors reviewed the manuscript.

Correspondence to Martin Hohmann.

The authors declare no competing interests.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

Hohmann, M., Kühn, D., Ni, D. et al. Relevant parameters for laser surgery of soft tissue. Sci Rep 14, 1263 (2024). https://doi.org/10.1038/s41598-024-51449-1

Download citation

Received: 21 September 2023

Accepted: 04 January 2024

Published: 13 January 2024

DOI: https://doi.org/10.1038/s41598-024-51449-1

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative